The UHPLC column used for isolating steroids/secosteroids is Waters ACQUITY UPLC HSS T3 Column (100Å, 1.8 µm, 2.1 mm × 100 mm, Part# 186003539).

Flow rate

When setting up an LC-MS/MS run, make sure that TSQ Altis (Figure 2) is correctly informed of the LC flow rate (Figure 1).

image
Figure 1. LC flow rate in Xcalibur Instrument Setup (pump settings).


image
Figure 2. LC flow rate fed to TSQ Altis (ion source properties) in Xcalibur Instrument Setup (NOT in FreeStyle).

Vanquish pump

When an LC-MS/MS sequence finishes running, the vanquish pump should be stopped manually: go to "Direct Control", click "Take Control" at the top, and uncheck the "Flow" box under the "Vanquish Pump" tab.

image
Figure 3. Stopping the vanquish pump.